Trial Purpose:
To evaluate three additional products on second supplied soil.
Date Run:
05/07/2004Experiment Procedure:
Three products were selected based on success on the fourth supplied contaminant. All were used at full strength in a 250 ml beaker at room temperature. Nine preweighed silver plated copper coupons were coated with the Houghton Chemical Cut-Max 135 (64742-52-5, 64741-96-4) using a hand held swab. Coupons were weighed a second time to determine the amount of soil added to each coupon. Three coupons were cleaned in each solution for 5 minutes using stir-bar agitation. After cleaning parts were weighed a final time and efficiencies were calculated.
Trial Results:
All three products removed over 99% of the oil in five minutes. The table lists the amount of soil added, the amount remaining and the efficiency for each coupon.
Cleaner | Initial wt | Final wt | % Removed |
OS 10 | 0.0880 | 0.0004 | 99.55 |
0.1503 | -0.0001 | 100.07 | |
0.1211 | 0.0007 | 99.42 | |
OS 20 | 0.1031 | -0.0001 | 100.10 |
0.1743 | 0.0005 | 99.71 | |
0.1313 | 0.0002 | 99.85 | |
OS 30 | 0.1355 | 0.0014 | 98.97 |
0.1048 | 0.0008 | 99.24 | |
0.1512 | 0.0012 | 99.21 |
Success Rating:
A follow up test, usually based on company input.Conclusion:
The same three products will be tested on the third supplied soil under the same conditions.