Browse past lab clients by general industry sectors
Trial Purpose:
To evaluate effectiveness of the provided cleaners in regards to removing synthetic blood from stainless steel.
Date Run:
02/05/2018Experiment Procedure:
Twelve pre-weighed stainless-steel coupons were soiled with synthetic blood using a swab and allowed to air dry for 24 hours before recording dirty weights. Two additional coupons were weighed but not soiled to provide a control. Two cleaners were prepared using the specified dilutions and temperature: Biogone (RTU at 22 C); Biogone (1:1 at 22 C).
Six coupons per cleaner were immersed in a beaker for two minutes. Coupons were rinsed in a de-ionized water bath for one minute and dried for 28 hours before recording clean weights.
Trial Results:
Both dilutions of Biogone were effective at removing the synthetic blood from the stainless-steel coupons.
| Cleaner | Initial Wt | Final Wt | %Removal | Average |
| Biogone RTU | 0.0253 | 0.0017 | 93.28 | 89.09 |
| 0.0226 | 0.0004 | 98.23 | ||
| 0.0183 | 0.0020 | 89.07 | ||
| 0.0257 | 0.0050 | 80.54 | ||
| 0.0201 | 0.0031 | 84.58 | ||
| 0.0215 | 0.0024 | 88.84 | ||
| 0.0066 | 0.0000 | - | ||
| Biogone 1:1 | 0.0208 | 0.0003 | 98.56 | 88.29
|
| 0.0258 | 0.0043 | 83.33 | ||
| 0.0264 | 0.0007 | 97.35 | ||
| 0.0254 | 0.0028 | 88.98 | ||
| 0.0259 | 0.0054 | 79.15 | ||
| 0.0204 | 0.0036 | 82.35 | ||
| 0.0066 | 0.0000 | - | ||
| Biogone 1:2 | 0.0184 | 0.0043 | 76.63 | 75.81 |
| 0.0148 | 0.0036 | 75.68 | ||
| 0.0213 | 0.0053 | 75.12 |
Success Rating:
A cleanliness study, addressing only various analytical techniques.Conclusion:
Both dilutions of Biogone were effective, as an overall average removal of 85% or more is considered effective. The RTU dilution was about 1% more effective than the 1:1 dilution. The product diluted 1:2 removed slightly more than 65%. The results were verified with visual observations.