Trial Purpose:
To evaluate the effectiveness of the provided cleaners in regards to removing synthetic blood from stainless steel.
Date Run:
03/06/2023Experiment Procedure:
36 pre-weighed stainless-steel coupons were coated with synthetic blood using a swab and were allowed to air dry (68 F) for 24 hours. One set of 18 coupons was set aside for the cleaner SuperNova 0.1 NpHD Neutral Detergent and the other set of 18 coupons was set aside for SuperNova 0.1 Multi-Enzymatic cleaner. Dirty weights were then recorded. The contaminated coupons were immersed in their respective cleaner in sets of three at a temperature of 160 F for SuperNova NpHD for two minutes and 140 F for SuperNova Multi-Enzymatic for two minutes. Following immersion, coupons were rinsed three times in a de-ionized water bath for one minute and dried for 24 hours before the final weights were recorded.
Trial Results:
Table 1: SuperNova 0.1 NpHD and 0.1 Multi-Enzymatic Cleaner Results
| Cleaner | Initial wt of cont. | Final wt of cont. | %Cont Removed | Average % Removal | Overall Average % Removal |
| SuperNova 0.1 NpHD Neutral Detergent-CSNB01G | 0.0064 | 0.0023 | 64.06 | 60.11 | 74.77 |
| 0.0038 | 0.0018 | 52.63 | |||
| 0.0011 | 0.0004 | 63.64 | |||
| 0.0030 | 0.0008 | 73.33 | 67.52 | ||
| 0.0047 | 0.0021 | 55.32 | |||
| 0.0023 | 0.0006 | 73.91 | |||
| 0.0058 | 0.0006 | 89.66 | 88.87 | ||
| 0.0031 | 0.0007 | 77.42 | |||
| 0.0218 | 0.0001 | 99.54 | |||
| 0.0030 | 0.0001 | 96.67 | 86.15 | ||
| 0.0093 | 0.0031 | 66.67 | |||
| 0.0143 | 0.0007 | 95.10 | |||
| 0.0044 | 0.0002 | 95.45 | 78.00 | ||
| 0.0046 | 0.0015 | 67.39 | |||
| 0.0052 | 0.0015 | 71.15 | |||
| 0.0038 | 0.0013 | 65.79 | 67.96 | ||
| 0.0063 | 0.0015 | 76.19 | |||
| 0.0042 | 0.0016 | 61.90 | |||
| SuperNova 0.1 Multi-Enzymatic Cleaner-CSNC01G | 0.0086 | 0.0010 | 88.37 | 75.97 | 81.97 |
| 0.0055 | 0.0017 | 69.09 | |||
| 0.0044 | 0.0013 | 70.45 | |||
| 0.0042 | 0.0008 | 80.95 | 81.72 | ||
| 0.0038 | 0.0006 | 84.21 | |||
| 0.0040 | 0.0008 | 80.00 | |||
| 0.0026 | 0.0004 | 84.62 | 81.47 | ||
| 0.0034 | 0.0008 | 76.47 | |||
| 0.0042 | 0.0007 | 83.33 | |||
| 1.0099 | 0.0072 | 99.29 | 77.38 | ||
| 0.0019 | 0.0007 | 63.16 | |||
| 0.0033 | 0.0010 | 69.70 | |||
| 0.0070 | 0.0007 | 90.00 | 88.37 | ||
| 0.0029 | 0.0004 | 86.21 | |||
| 0.0036 | 0.0004 | 88.89 | |||
| 0.0042 | 0.0006 | 85.71 | 86.92 | ||
| 0.0050 | 0.0011 | 78.00 | |||
| 0.0034 | 0.0001 | 97.06 |
While SuperNova 0.1 NpHD Cleaner had some high contaminant removal percentages, SuperNova 0.1 Multi-Enzymatic Cleaner was overall more consistent with its average removal of the synthetic blood on stainless steel.
Success Rating:
Results successful using TACT (time, agitation, concentration, and temperature, as well as rinsing and drying) and/or other cleaning chemistries examined.Conclusion:
SuperNova 0.1 Multi-Enzymatic Cleaner was more effective at removing synthetic blood from stainless steel with an overall average removal of 81.97%. SuperNova 0.1 NpHD Neutral Detergent was slightly less effective with an overall average removal of 74.77%.