Trial #0

Trial #0

To evaluate effectiveness of the provided cleaners in regards to removing synthetic blood from stainless steel.

2018

25

1

5

0

0

Coupon

02/02/2018

2.00

Immersion/Soak

James Keats
Justin Rainaud

Stainless Steel

Blood

DI water rinse

Air dry

Gravimetric

Twelve pre-weighed stainless-steel coupons were soiled with synthetic blood using a swab and allowed to air dry for 24 hours before recording dirty weights. Two additional coupons were weighed but not soiled to provide a control.  Two cleaners were prepared using the specified dilutions and temperature: Biogone (RTU at 22 C); Biogone (1:1 at 22 C).

 Six coupons per cleaner were immersed in a beaker for two minutes.  Coupons were rinsed in a DI water bath and dried for 48 hours before recording clean weights.

Cleaner Initial Wt Final Wt %Removal Average

 

 

 

Biogone RTU

0.0020 0.0005 75.00

 

 

 

65.07

0.0029 0.0013 55.17
0.0101 0.0015 85.15
0.0143 0.0009 93.71
0.0037 0.0023 37.84
0.0062 0.0035 43.55
0.0046 0.0003 - Control

 

 

 

Biogone 1:1

0.0067 0.0035 47.76

 

 

 

40.23

 

0.0089 0.0043 51.69
0.0132 0.0082 37.88
0.0075 0.0033 56.00
0.0092 0.0092 0.00
0.0052 0.0027 48.08
0.0019 0.0008 - Control

Both dilutions of Biogone were not effective.  The RTU dilution was about 25% more effective than the 1:1 dilution.  The same ranking was observed using visual analysis. 