Trial #1

Trial #1

To evaluate effectiveness of the provided cleaners in regards to removing synthetic blood from stainless steel.

2018

25

5

0

0

Part

02/05/2018

2.00

Immersion/Soak

James Keats
Vinh Tran

Stainless Steel

Blood

DI water rinse

Air dry

Gravimetric

Twelve pre-weighed stainless-steel coupons were soiled with synthetic blood using a swab and allowed to air dry for 24 hours before recording dirty weights. Two additional coupons were weighed but not soiled to provide a control.  Two cleaners were prepared using the specified dilutions and temperature: Biogone (RTU at 22 C); Biogone (1:1 at 22 C).

Six coupons per cleaner were immersed in a beaker for two minutes. Coupons were rinsed in a de-ionized water bath for one minute and dried for 28 hours before recording clean weights.

Both dilutions of Biogone were effective at removing the synthetic blood from the stainless-steel coupons.

Cleaner Initial Wt Final Wt %Removal Average
Biogone RTU 0.0253 0.0017 93.28 89.09
0.0226 0.0004 98.23
0.0183 0.0020 89.07
0.0257 0.0050 80.54
0.0201 0.0031 84.58
0.0215 0.0024 88.84
0.0066 0.0000 -  
Biogone 1:1 0.0208 0.0003 98.56

 88.29

 

0.0258 0.0043 83.33
0.0264 0.0007 97.35
0.0254 0.0028 88.98
0.0259 0.0054 79.15
0.0204 0.0036 82.35
0.0066 0.0000 -  
Biogone 1:2 0.0184 0.0043 76.63 75.81
0.0148 0.0036 75.68
0.0213 0.0053 75.12

Both dilutions of Biogone were effective, as an overall average removal of 85% or more is considered effective. The RTU dilution was about 1% more effective than the 1:1 dilution.  The product diluted 1:2 removed slightly more than 65%. The results were verified with visual observations. 