Trial #2

Trial #2

To evaluate effectiveness of the provided cleaner in regards to removing synthetic blood from stainless steel.

2018

26

5

0

0

Coupon

09/15/2018

2.00

Immersion/Soak

Sabrina Apel

Stainless Steel

Blood

DI water rinse

Air dry

Gravimetric

Three pre-weighed stainless-steel coupons were coated with synthetic blood using a swab and allowed to air dry (68 F) for 24 hours. Dirty weights were recorded before immersing coupons into the RTU cleaner provided by the company at room temperature (68 F) for two minutes. Coupons were rinsed three times in a de-ionized water bath for one minute and dried for 24 hours before recording the final weights.

Cleaner Substrate Initial wt. of cont. Final wt. of cont. %Cont. Removed % Average
Biogone Disinfectant RTU Stainless Steel 0.0046 0.0012 73.91 75.74
0.0039 0.0008 79.49
0.0042 0.0011 73.81

The cleaner, Biogone Disinfectant RTU, was ineffective in removing synthetic blood from stainless steel coupons with an overall average removal of 76%. The results were verified with visual observations.