Trial #3

Trial #3

To evaluate effectiveness of the provided cleaner in regards to removing synthetic blood from stainless steel.

2019

26

5

0

0

Coupon

01/03/2019

2.00

Immersion/Soak

Phillip Demers

Stainless Steel

Blood

DI water rinse

Air dry

Gravimetric

Three pre-weighted stainless-steel coupons were soiled with synthetic blood using a swab and allowed to air dry at room temperature for 24 hours before recording the dirty weights. The coupons were immersed in a beaker with the cleaner for two minutes. Coupons were rinsed in a de-ionized water bath for one minute and dried for 24 hours before recording clean weights.

Cleaner Initial Wt. Final Wt. % Removal Average
BioGone RTU 0.0010 0.0006 40.00 47.78
0.0016 0.0008 50.00
0.0015 0.0007 53.33

BioGone RTU was not considered effective in removing synthetic blood from stainless steel.  

No relation

Name Class Section
Document Evaluation #0 Evaluation 3