Trial #1
Evaluate the efficacy of Client's cleaner formulations on the removal of synthetic blood soil from a stainless steel substrate.
36
0
1
23
Coupon
02/10/2026
0.00
Immersion/Soak
Alexander Symko Amelia Wagner Tatyanna Moreland Junior
Stainless Steel
Blood
DI water rinse
Air dry
Gravimetric
15 pre-weighed stainless-steel coupons were coated with synthetic blood using a swab and were allowed to air dry (68 F) for 24 hours. The supplied cleaner concentrates were then diluted by a ratio of 0.5 oz/ gallon of water. Dirty weights were then recorded. The contaminated coupons were immersed in their respective cleaner at room temperature for two minutes. Following immersion, coupons were rinsed three times in a de-ionized water bath for one minute and dried for 24 hours before the final weights were recorded.
Based on the results of this report, Caseprep and Supernova 1.0 were the most effective at removing the blood soil from the substrate, but all cleaners attained a high level of cleanliness (>80% removal) and as such can all be considered effective in the cleaning application.