Trial #3
Five sets of 26-28 titanium parts were provided pre-contaminated with J2. Cleaners were heated to vendor recommended temperatures before immersing the parts into an ultrasonic tank for 15 minutes. Visual observations were recorded overall with notes and a set of three parts per cleaner were evaluated by three lab technicians, labelled A-C in Table 1, for cleaner and contaminant residue according using the rating key below.
Five sets of 26-28 titanium parts were provided pre-contaminated with J2. Cleaners were heated to vendor recommended temperatures before immersing the parts into an ultrasonic tank for 15 minutes. Visual observations were recorded overall with notes and a set of three parts per cleaner were evaluated by three lab technicians, labelled A-C in Table 1, for cleaner and contaminant residue according using the rating key below.
Rating Key:
| Rating Key
|
| 1
|
No Residue
|
| 2
|
Slight Residue
|
| 3
|
Noticeable Residue
|
| 4
|
Considerable Residue
|
| 5
|
Severe Residue
|
Table 1: Visual Rating Results
| Cleaner
|
Conc. (%)
|
Temp. (°F)
|
Before Cleaning Rating
|
After Cleaning Rating
|
Average Score
|
| A
|
B
|
C
|
A
|
B
|
C
|
| Dowanol PnBGE
|
100
|
100
|
4
|
5
|
4
|
2
|
2
|
1.5
|
2
|
| Ozzy Juice SW 3
|
100
|
110
|
3
|
4
|
2.5
|
3
|
| Metalnox M6386
|
100
|
130
|
1
|
1
|
1.5
|
1
|
| Sta-Sol ESS 160
|
100
|
106
|
2
|
2
|
1.5
|
2
|
| Liquinox
|
1
|
150
|
5
|
5
|
5
|
5
|
Visual Observations:
Dowanol PnBGE - Slight cleaner residue and soil residue that remained on parts after cleaning.
Ozzy Juice SW3 - Slight cleaner residue and soil residue that remained on parts after cleaning.
Metalnox M6386 - No cleaner residue or soil residue on parts after cleaning
Sta-Sol ESS-160 - No cleaner residue or soil residue on parts after cleaning, however, it took a while for the parts to completely dry
Liquinox - Cleaner residue and soil residue remained on parts after cleaning. Cleaner will need a rinse step going forward with testing.
Dowanol PnBGE, Metalnox 6386, and Sta-Sol ESS 160 were the most effective at removing J2 soil from titanium parts using heated ultrasonics.