Trial Purpose:
To evaluate aqueous alternatives using ultrasonic cleaning
Date Run:
08/07/2008Experiment Procedure:
Six products from previous trials were selected for testing. Five of the products were diluted to 20% and the sixth product was diluted to 5%. Products were heated to 130 F and immersed in a Branson 40 kHz ultrasonic tank and degassed for at least five minutes.
Eighteen preweighed coupons were coated with the supplied cutting fluid using a handheld swab and weighed a second time to determine the amount of soil added. Three coupons were cleaned with ultrasonic energy for five minutes, rinsed in tap water for 15 seconds at 120 F and dried using compressed air at room temperature for 30 seconds. Final weights were recorded, and efficiencies calculated for each coupon cleaned.
Trial Results:
Five of the six products removed more than 85% of the oil in the five minutes of cleaning. The last product removed 80%. The table lists the amount of soil added, the amount remaining and the efficiency for each coupon cleaned.
Cleaner | Initial wt | Final wt | % Removed |
Shopmaster LpH | 0.1368 | 0.0234 | 82.89 |
0.2761 | 0.0159 | 94.24 | |
0.2485 | 0.0171 | 93.12 | |
Aquavantage 1400 GD | 0.2711 | 0.0024 | 99.11 |
0.3318 | 0.0013 | 99.61 | |
0.4337 | 0.0017 | 99.61 | |
Shopmaster HP | 0.3196 | 0.0404 | 87.36 |
0.3563 | 0.0439 | 87.68 | |
0.3455 | 0.0447 | 87.06 | |
Inproclean 3800 | 0.4706 | 0.0020 | 99.58 |
0.3914 | -0.0041 | 101.05 | |
0.3043 | -0.0008 | 100.26 | |
Bio T 300 B | 0.2473 | 0.0597 | 75.86 |
0.3007 | 0.0518 | 82.77 | |
0.3603 | 0.0604 | 83.24 | |
SC Aircraft & Metal Cleaner | 0.4117 | 0.0107 | 97.40 |
0.2978 | 0.0547 | 81.63 | |
1.3744 | 0.0217 | 98.42 |
Success Rating:
Results successful using TACT (time, agitation, concentration, and temperature, as well as rinsing and drying) and/or other cleaning chemistries examined.Conclusion:
The same products will be evaluated on the second supplied contaminant following the same conditions.