Trial Number 1

Trial Purpose:

To evaluate the effectiveness of aqueous-based and solvent-based cleaners at the removal of American Lapping Compound by S.E Shires using heated immersion on copper, brass, and copper/nickel (90/10) substrates by gravimetric and visual analysis.

Date Run:

03/27/2020

Experiment Procedure:

Three pre-weighed coupons per substrate were soiled with American Lapping Compound, using a cotton swab. Each set of coupons were immersed into a beaker with one of the six cleaners for 30 minutes at the recommended temperature. During this process, any noticeable soil removal was observed at five minute intervals. After immersing, final weights were recorded and efficiency was calculated for each coupon cleaned.

Trial Results:

The soil visually was not removed from each substrate for each cleaner. After drying for 24 hours, some coupons were still wet from the cleaner, indicating the soil may have absorbed moisture causing a negative percent contaminant removal.

 

Cleaner Temperature (⁰F) Substrate Initial wt of cont. Final wt of cont. %Cont Removed % Average Overall % Removal
Aquavantage 1400 10% 130 Copper 0.5628 0.5622 0.11 0.07 -0.46
0.4427 0.4425 0.05
0.6854 0.6851 0.04
Brass 0.5269 0.5269 0.00 -0.03
0.6078 0.6078 0.00
0.6471 0.6476 -0.08
Copper/Nickel 0.2655 0.2740 -3.20 -1.43
0.3271 0.3288 -0.52
0.2418 0.2432 -0.58
Surface Cleanse 930 5% 130 Copper 0.3485 0.3946 -13.23 -10.78 -10.23
0.4546 0.4951 -8.91
0.3292 0.3628 -10.21
Brass 0.4083 0.4505 -10.34 -7.52
0.4678 0.4887 -4.47
0.3742 0.4032 -7.75
Copper/Nickel 0.2624 0.2970 -13.19 -12.38
0.2464 0.2949 -19.68
0.2720 0.2836 -4.26
SC Aircraft & Metal Cleaner 5% 130 Copper 0.4271 0.4263 0.19 0.45 0.77
0.5026 0.5014 0.24
0.4486 0.4445 0.91
Brass 0.3207 0.3198 0.28 0.32
0.3558 0.3545 0.37
0.7226 0.7204 0.30
Copper/Nickel 0.5181 0.4963 4.21 1.55
0.2987 0.2981 0.20
0.3003 0.2996 0.23
Micro 90 2% 130 Copper 0.4578 0.4535 0.94 1.38 2.10
0.4464 0.4401 1.41
0.3718 0.3651 1.80
Brass 0.3912 0.3855 1.46 2.47
0.4484 0.4396 1.96
0.5194 0.4987 3.99
Copper/Nickel 0.3278 0.3245 1.01 2.46
0.2653 0.2606 1.77
0.2832 0.2702 4.59
LF2100 5% 130 Copper 0.4916 0.4887 0.59 -4.88 -2.65
0.6743 0.4938 26.77
0.4232 0.6009 -41.99
Brass 0.5039 0.6260 -24.23 -3.08
0.6477 0.5730 11.53
0.4210 0.4064 3.47
Copper/Nickel 0.3576 0.3619 -1.20 0.02
0.4790 0.4805 -0.31
0.2877 0.2832 1.56
Liquinox 130 Copper 0.3873 0.3808 1.68 4.38 3.48
0.3714 0.3361 9.50
0.6031 0.5913 1.96
Brass 0.5064 0.5011 1.05 3.05
0.4938 0.4810 2.59
0.4466 0.4220 5.51
Copper/Nickel 0.4393 0.4365 0.64 3.01
0.3918 0.3904 0.36
0.2599 0.2390 8.04

Success Rating:

Test showed little or no promise.

Conclusion:

Each cleaner was not effective in removing American Lapping Compound contaminant from copper, brass, and copper/nickel (90/10) coupons. Next step is to only test only on copper coupons with a higher dilution unheated based on vendor recommendation before adding ultrasonics. 