Trial Purpose:

To evaluate effectiveness of the provided cleaner in regards to removing synthetic blood from stainless steel.

Date Run:

09/16/2018

Experiment Procedure:

Three pre-weighed stainless-steel coupons were coated with synthetic blood using a swab and allowed to air dry (68 F) for 24 hours. Dirty weights were recorded before immersing coupons into the RTU cleaner provided by the company at room temperature (68 F) for two minutes. Coupons were rinsed three times in a de-ionized water bath for one minute and dried for 24 hours before recording the final weights.

Trial Results:

Cleaner Substrate Initial wt. of cont. Final wt. of cont. %Cont. Removed % Average
Biogone Disinfectant RTU Stainless Steel 0.0046 0.0012 73.91 75.74
0.0039 0.0008 79.49
0.0042 0.0011 73.81

Success Rating:

A cleanliness study, addressing only various analytical techniques.

Conclusion:

The cleaner, Biogone Disinfectant RTU, was ineffective in removing synthetic blood from stainless steel coupons with an overall average removal of 76%. The results were verified with visual observations.